Titre du document / Document title
Laboratory assessment of a user-assembled oxygen generation kit
Auteur(s) / Author(s)
BHIMSAN N. R. ;
ROUT C. C. ;
MURRAY W. B. ;
Affiliation(s) du ou des auteurs / Author(s) Affiliation(s)
Univ. Natal, fac. medicine, dep. anaesthetics, Durban, AFRIQUE DU SUD
Résumé / Abstract
A laboratory assessment was made of a user-assembled oxygen generation it designed for use in medical emergencies in the community. Twenty-five kits were studied at five different water temperatures and the rate of oxygen generation and its concentration were measured. The apparatus produced a mean (SD) oxygen volume of 37 (1.98) 1 which was not affected by the starting water temperature. The mean (SD) rate of flow varied from 2.2 (0.04) l.min
-1 at 16
oC to 10.3 (2.74) l.min
-1 at 40
oC but the duration of flow varied inversely. Oxygenation concentrations ranging from 30-51% were produced at starting water temperatures from 22
oC-40
oC, respectively
Revue / Journal Title
Anaesthesia
ISSN 0003-2409
CODEN ANASAB
Source / Source
1994, vol. 49, n
o5, pp. 419-421 (3 ref.)
Langue / Language
Anglais
Editeur / Publisher
Blackwell, Oxford, ROYAUME-UNI
(1946)
(Revue)
Mots-clés anglais / English Keywords
Generator ;
Oxygen ;
Apparatus ;
Kit ;
Performance evaluation ;
Mots-clés français / French Keywords
Générateur ;
Oxygène ;
Appareil ;
Kit ;
Evaluation performance ;
Mots-clés espagnols / Spanish Keywords
Generador ;
Oxígeno ;
Aparato ;
Kit ;
Evaluación prestación ;
Localisation / Location
INIST-CNRS, Cote INIST : 7599, 35400004563137.0140
Nº notice refdoc (ud4) : 4067534