Résumé / Abstract
A novel electrochemical immunosensor based on the principles of liposome signal amplification and immunochromatography has been developed for the determination of theophylline. The immunosensor is composed of two major parts including a screen-printed electrode and an immunochromatographic nitrocellulose membrane strip. On the membrane, anti-theophylline antibody is immobilized in an antibody competition zone and hexacyanoferrate(II))-loaded liposomes are immobilized in a signal generation zone. When a theophylline sample solution is applied to the immunosensor pre-loaded with theophylline-melittin conjugate in sample loading zone, the theophylline and theophylline-melittin conjugate migrate through the anti-theophylline antibody zone, where competitive binding occurs. Unbound theophylline-melittin conjugate further migrates into the signal generation zone, where it disrupts the liposomes to release the electroactive hexacyanoferrate(II) which is then detected amperometrically. The current produced is directly proportional to the concentration of hexacyanoferrate(II) which, in turn, can be related to the concentration of free analyte in the sample. The detection limit of 5 μg ml
-1 enables the present immunosensor to be used to monitor theophylline over the clinically relevant ranges (10-20 μg ml
-1) with a one-step assay within 20 min.
Revue / Journal Title
Analytica chimica acta
ISSN 0003-2670
CODEN ACACAM
Source / Source
1999, vol. 380, n
o1, pp. 17-26 (34 ref.)
Langue / Language
Anglais
Editeur / Publisher
Elsevier, Amsterdam, PAYS-BAS
(1947)
(Revue)
Mots-clés anglais / English Keywords
Immunosensor ;
Liposome ;
Electrodes ;
Membrane ;
Immobilization ;
Antibody ;
Performance evaluation ;
Theophylline ;
Quantitative analysis ;
Biosensor ;
Portable ;
Electrochemistry ;
Mots-clés français / French Keywords
Immunodétecteur ;
Liposome ;
Electrode ;
Membrane ;
Immobilisation ;
Anticorps ;
Evaluation performance ;
Théophylline ;
Analyse quantitative ;
Biodétecteur ;
Portatif ;
Electrochimie ;
Immunochromatographie ;
Mots-clés espagnols / Spanish Keywords
Inmunodetector ;
Liposoma ;
Electrodo ;
Membrana ;
Inmovilización ;
Anticuerpo ;
Evaluación prestación ;
Teofilina ;
Análisis cuantitativo ;
Biodetector ;
Portátil ;
Electroquímica ;
Localisation / Location
INIST-CNRS, Cote INIST : 3950, 35400007363097.0030
Nº notice refdoc (ud4) : 1680387