Clinical comparison of the ocular blood flow tonograph and the Goldmann applanation tonometer
Auteur(s) / Author(s)
ESGIN H.
(1) ;
ALIMGIL M. L.
(1) ;
ERDA S.
(1) ;
Affiliation(s) du ou des auteurs / Author(s) Affiliation(s)
(1) Department of Ophthalmology, University of Trakya, Edirne, TURQUIE
Résumé / Abstract
Purpose. To compare the IOP readings of an ocular blood flow (OBF) tonograph (OBF Labs UK Ltd) with Goldmann applanation tonometry. Methods. 194 patients were studied. In group 1, the software version 8.2 of the OBF tonograph was used in 214 eyes of 107 patients and in group 2, the 11.2 version of the tonograph was used in 174 eyes of 87 patients. Results. We found 63% of the OBF tonograph readings to be within ±2mm Hg of the Goldmann applanation tonometry readings in group 1 and 60% in group 2. In group 1 the correlation coefficient between the readings of the two instruments was 0. 71 and 0. 82 in group 2. The mean value for the paired differences in group 1 was-1.34±2.75mm Hg overall and -1.04±2.91 mm Hg overall in group 2. There was a significant difference between the Goldmann applanation tonometer and OBF tonograph readings in the 8-10mm Hg interval (p<0.001) in group 1, but not in group 2 (p>0.1). Conclusions. The 11.2 version of the OBF tonograph is more accurate than the 8.2 version and measures IOP in a manner that corresponds well to the Goldmann applanation tonometer in the 8-10mm Hg and 21-29mm Hg intervals. It corresponds fairly well in the 11-20mm Hg interval.
Revue / Journal Title
European Journal of Ophthalmology
ISSN 1120-6721
Source / Source
1998, vol. 8, n
o3, pp. 162-166 (7 ref.)
Langue / Language
Anglais
Editeur / Publisher
Wichtig, Milano, ITALIE
(1991)
(Revue)
Mots-clés anglais / English Keywords
Glaucoma (eye) ;
Measurement method ;
Intraocular pressure ;
Tonography ;
Blood flow ;
Eye ;
Tonometry ;
Measurement accuracy ;
Comparative study ;
Human ;
Eye disease ;
Mots-clés français / French Keywords
Glaucome ;
Méthode mesure ;
Tonus oculaire ;
Tonographie ;
Débit sanguin ;
Oeil ;
Tonométrie ;
Précision mesure ;
Etude comparative ;
Homme ;
Oeil pathologie ;
Mots-clés espagnols / Spanish Keywords
Glaucoma (ojo) ;
Método medida ;
Tono ocular ;
Tonografía ;
Flujo sanguíneo ;
Ojo ;
Tonometría ;
Precisión medida ;
Estudio comparativo ;
Hombre ;
Ojo patología ;
Localisation / Location
INIST-CNRS, Cote INIST : 22516, 35400007301279.0080
Nº notice refdoc (ud4) : 1608602